Tischabzug

Bench-type fume cupboard

All-rounders for general use

EXPLORIS® bench-type fume cupboards for general use can be used universally in the laboratory. They are ideal for dealing with organic solvents and are also suitable for working with cold or diluted acids. Depending on the room height of the laboratory, there is a choice between a high and a low design.

Learn more

Properties and benefits

  • Certified according to DIN EN 14175 Part 2 and Part 3
  • Low operating costs due to low airflow
  • Extra large interior depth due to special worktop shape

Equipment options

  • Sash options
    • one-piece sash (fume cupboards with standard front height)
    • telescopic sash (fume cupboards with reduced front height)
    • two-piece laminated safety glass (with and without sliding glass panels)
  • Foot switch for operating the sash
  • Interior: coated steel, optional lining in Trespa® Toplab®VERTICAL
  • Drawer with full pull-out, cushioned automatic closure and label
  • Worktops in different materials
  • LED interior lighting
  • Programmable sockets in the fume cupboard chamber
  • Individual gas, water and electrical fittings with qualtity products from BROEN – LAB A/S, Spectron Gas Control Systems GmbH and GCE GmbH
  • Integrated fuses / RCCBs
  • Scaffolding parts for the installation on the rear panel of the fume cupboard
  • Cable entry through the inlet profile on the worktop
  • Cable bushing in the fume cupboard side
  • Window in the fume cupboard side
  • Also available as fume cupboard for low ceiling height with reduced front height (2300mm) and floor-mounted service columns

Technical details

Dimensions EXPLORIS® Bench-type fume cupboard high / low design

Tischabzug Maße

EXPLORIS® Bench-type fume cupboards

Dimensions in mm / Volume flow in m³/h
Width 1200 1500 1800 2000 2100
Front height 2300/2550 2300/2550 2300/2550 2300/2550 2300/2550
Housing height 2300/2600 2300/2600 2300/2600 2300/2600 2300/2600
Depth 900 900 900 900 900
Usable space width 1100 1400 1700 1900 2000
Usable space height 1190/1490 1190/1490 1190/1490 1190/1490 1190/1490
Usable space depth 800 800 800 800 800
Working height 900 900 900 900 900
Minimum permissible volume flow (alarm value) 450 530 680 800 800
Volume flow to be adjusted 480 600 720 840 840
Maximum permissible volume flow 1500 1500 1500 1500 1500