Laboratory for research and contract work of the Institute for Environmental Technology and Energy Economics in Hamburg-Harburg.
Bench type fume cupboards
All-rounders for general use
Bench type fume cupboards
All-rounders for general use
Fume cupboard for low ceiling height with unique low installation height (2250 mm)
Above-average usable worktop depth
Underbench storage for tall utensils up to 720 mm
Metallic construction enables columns with hollow spaces at the front – for easy maintenance and protection against air leakage
Interior: coated steel, optional lining in Trespa®
Side or front service connections for easy tapping of water, gas, etc.
Fume cupboard for low ceiling height with unique low installation height (2250 mm)
Above-average usable worktop depth
Underbench storage for tall utensils up to 720 mm
Metallic construction enables columns with hollow spaces at the front – for easy maintenance and protection against air leakage
Interior: coated steel, optional lining in Trespa®
Side or front service connections for easy tapping of water, gas, etc.
Width | 1200 | 1500 | 1800 | 2000 | 2100 |
---|---|---|---|---|---|
Front height | 2300/2550 | 2300/2550 | 2300/2550 | 2300/2550 | 2300/2550 |
Chamber height | 2300/2600 | 2300/2600 | 2300/2600 | 2300/2600 | 2300/2600 |
Depth | 900 | 900 | 900 | 900 | 900 |
Usable space width | 1100 | 1400 | 1700 | 1900 | 2000 |
Usable space height | 1190/1490 | 1190/1490 | 1190/1490 | 1190/1490 | 1190/1490 |
Usable space depth | 800 | 800 | 800 | 800 | 800 |
Work height | 900 | 900 | 900 | 900 | 900 |
Minimal permissable volume flow rate* | 450 | 530 | 680 | 800 | 800 |
Specified volume flow rate | 480 | 600 | 720 | 840 | 840 |
Maximum permissible volume flow rate | 1500 | 1500 | 1500 | 1500 | 1500 |
* Alarm value
Your work is what matters
When selecting a fume cupboard, the most important considerations are the activities you do and the substances you use.
Fume cupboards for general use with cemical-resistant Trespa Toplab VERTICAL® lining
Product data
The certification according to DIN EN 14175 part 2 and 3 allows EXPLORIS® fume cupboards for general use to be used universally in the laboratory. Below you will find all information about dimensions and volume flow.
Fume cupboard control TouchTronic
Easy operation of fume cupboards
As a control element for a wide range of functions, the new TouchTronic combines fume cupboard control and monitoring with high computing power. The combination of work steps in one unit saves time and also protects everyone else in the lab: warnings can be easily shown on the display. Even when wearing gloves, the IPS touch screen recognises the touch.
Fume cupboard control TouchTronic
Easy operation of fume cupboards
As a control element for a wide range of functions, the new TouchTronic combines fume cupboard control and monitoring with high computing power. The combination of work steps in one unit saves time and also protects everyone else in the lab: warnings can be easily shown on the display. Even when wearing gloves, the IPS touch screen recognises the touch.